Simplex 600 und 650 Manual and pneumatic sampler
Simplex 600
Semi-automatic sampler in the Simplex 600 version with one sampler output, in manual and pneumatic versions.


Simplex 650
Semi-automatic sampler in the Simplex 650 version with one outlet and one cleaning inlet for cleaning and disinfecting the sampler, in manual and pneumatic versions.


Simplex 600/650 Technical data

Sampler:
Pneumatic and hand lever version
Seal:
Viton sealing cone, O-ring silicone (FDA approved)
Material:
1.4404 in contact with product; 1.4301 not in contact with product
Line pressure:
from 0 to 10 bar opening at 6 bar compressed air up to 16 bar with hand lever
Line pressure shock:
up to max. 40 bar without opening
Compressed air connection:
DN 6 quick-release coupling
Opening travel:
approx. 3 mm
Flow size:
DN 5 mm
Connection inlet:
Varivent connection from DN25/32 and DN40/150mm
Dampable:
up to 125°C
Tightness:
up to 18 bar with liquid medium, up to 8 bar with air/CO2/nitrogen
Weight:
1.2 kg
Inch-Id-No.:
8481 8019
Maintenance effort:
max. 2 seals, time required to change approx. 1-2 minutes
Maintenance interval:
depending on medium and cleaning interval, every 3 - 12 months
| Article-#: | AE00.1600.#.# / AE00.1650.#.# | ||
|---|---|---|---|
| Article-#: | AE00.1600.#input.#output / AE00.1650.#input.#output | ||
| # | Sampler input | # | Sampler output |
| .1 | Varivent cover Type N | .1 | Quick-release coupling DN5 |
| .2 | Varivent cover Type F | .2 | 1/2" Aseptik |
| .3 | Thread connection Bio-Con | .3 | 1/2" Standard |
| .4 | Tri-Clamp 1 1/2" | .4 | 5/8" Standard |
| .5 | Customer specification | ||
| .6 | 1/2" Cromer | ||
| .7 | 3/8" Standard | ||
AE00.1600.1.1
Sampler Simplex 600
Inlet connection Varivent cover type N,
sample outlet quick-coupling DN5
AE00.1650.1.1
Sampler Simplex 650
Inlet connection Varivent cover type N,
cleaning inlet and sample outlet
quick-release coupling DN5
Sensor for monitoring the open/closed switching status
Accessories